From WILHELM@sumop1.nascom.nasa.gov Tue Nov 18 16:45:30 2003 Date: Mon, 7 Jun 1999 17:30:56 GMT From: WILHELM@sumop1.nascom.nasa.gov Reply-To: sumer-soft@esa.nascom.nasa.gov To: sumer-soft@esa.nascom.nasa.gov Cc: WILHELM@sumop1.nascom.nasa.gov Subject: SUMER-SOFT: SUMER radiometry Subject: SUMER Radiometry Programme Dear colleagues, I am sorry to announce an error in the radiometry programme, which Udo Schuehle has just found. The error is in the new magnification programme and occurs for detector B and in second order only. The spectral resolution element came out twice as large as it should be. The new programme was put into the software tree on 26 November 1998 and was updated today with a correction of the error. This error had an effects on radiometry outputs only for spectral radiances (per Angstroem), but not for line radiances. So, if you have done any calibration work involving spectral radiances and detector B in second order recently, I am afraid you have to check your results. I am sorry for the inconvenience. With best regards, Klaus Wilhelm